×
近期发现有不法分子冒充我刊与作者联系,借此进行欺诈等不法行为,请广大作者加以鉴别,如遇诈骗行为,请第一时间与我刊编辑部联系确认(《中国物理C》(英文)编辑部电话:010-88235947,010-88236950),并作报警处理。
本刊再次郑重声明:
(1)本刊官方网址为cpc.ihep.ac.cn和https://iopscience.iop.org/journal/1674-1137
(2)本刊采编系统作者中心是投稿的唯一路径,该系统为ScholarOne远程稿件采编系统,仅在本刊投稿网网址(https://mc03.manuscriptcentral.com/cpc)设有登录入口。本刊不接受其他方式的投稿,如打印稿投稿、E-mail信箱投稿等,若以此种方式接收投稿均为假冒。
(3)所有投稿均需经过严格的同行评议、编辑加工后方可发表,本刊不存在所谓的“编辑部内部征稿”。如果有人以“编辑部内部人员”名义帮助作者发稿,并收取发表费用,均为假冒。
                  
《中国物理C》(英文)编辑部
2024年10月30日

Optimum condition of efficiency functions for HPGe γ-ray detectors in the 121-1408 keV energy range

Get Citation
CHEN Zhi-Lin, SONG Guo-Yang, MU Long, WANG He-Yi, XING Shi-Xiong, GUO Hong-Bo, LIAO Zhen-Xing, CHEN Ping and HUA Sheng. Optimum condition of efficiency functions for HPGe γ-ray detectors in the 121-1408 keV energy range[J]. Chinese Physics C, 2010, 34(5): 560-564. doi: 10.1088/1674-1137/34/5/009
CHEN Zhi-Lin, SONG Guo-Yang, MU Long, WANG He-Yi, XING Shi-Xiong, GUO Hong-Bo, LIAO Zhen-Xing, CHEN Ping and HUA Sheng. Optimum condition of efficiency functions for HPGe γ-ray detectors in the 121-1408 keV energy range[J]. Chinese Physics C, 2010, 34(5): 560-564.  doi: 10.1088/1674-1137/34/5/009 shu
Milestone
Received: 2009-06-28
Revised: 2009-11-26
Article Metric

Article Views(2670)
PDF Downloads(487)
Cited by(0)
Policy on re-use
To reuse of subscription content published by CPC, the users need to request permission from CPC, unless the content was published under an Open Access license which automatically permits that type of reuse.
通讯作者: 陈斌, [email protected]
  • 1. 

    沈阳化工大学材料科学与工程学院 沈阳 110142

  1. 本站搜索
  2. 百度学术搜索
  3. 万方数据库搜索
  4. CNKI搜索

Email This Article

Title:
Email:

Optimum condition of efficiency functions for HPGe γ-ray detectors in the 121-1408 keV energy range

    Corresponding author: CHEN Zhi-Lin,

Abstract: 

The optimum condition of three commonly used functions in the Genie 2000 γ spectra analysis software have been studied in the 121—1408 keV energy range. The three functions are applied for fitting the full-energy peak efficiency of the HPGe gamma-ray detector. A detailed procedure to obtain the optimum condition is described. The HPGe detector is calibrated at 11 cm by three radioactive sources of point form (152Eu, 137Cs, 60Co) providing 11 energy peaks. After data processing, results shows that the three functions used in the Genie 2000 gamma spectra analysis software fit best at orders 3—5. Lastly the standard radioactive source 133Ba is chosen to validate the results. Differences between the standard activity of 133Ba and the result obtained from the fitting functions are below 1.5%. Therefore the optimum orders of the three functions used in the Genie 2000 γ spectra analysis software are 3—5 with the 11 energy peaks.

    HTML

Reference (1)

目录

/

DownLoad:  Full-Size Img  PowerPoint
Return
Return